As the art and science of infrared thermography has matured, infrared test equipment has evolved. Technological advances during the past 15 years have enabled equipment manufacturers to design infrared imagers capable of providing real-time noncontact temperature measurements. The evolution of modern imaging radiometers has progressed to the point that most commercially available infrared cameras are now capable of providing noncontact temperature measurements.
When using a radiometer to measure object temperatures, several important factors must be addressed. Among these are target emittance, atmospheric attenuation, spectral response of the imaging radiometer, and measurement spot size. While spot size is not defined in currently published standards, it is generally defined as the area from which radiometric or temperature data are derived.
For accurate temperature measurement, the spot size of the radiometer must be smaller than the target being measured. Should the spot size be larger than the target, error will be introduced into the measurement. The amount of error will be dependent upon a number of factors, none of which can be corrected for by any means including radiometric software.
Lack of spot size information
While many nonimaging radiometers now employ a laser sighting system to project relative spot size onto the object being measured, no such sighting system exists for imaging radiometers. As such, it is impossible for a thermographer to visually gauge the spot size of the imager in use. In fact, many thermographers mistakenly believe that imaging radiometers are capable of producing accurate temperatures equal to the pixel size of the imager.
Although many modern infrared imagers are capable of measuring temperature, infrared equipment manufacturers generally do not provide information regarding spot size or how to calculate it. In contrast, nonimaging radiometers have always expressed spot size as a distance to target ratio such as 50:1. Applying this ratio, at 50 in. from a target, the spot size would be 1 in. This methodology provides a simple and quick means for calculating spot size at any distance from the target.
To date, manufacturers of infra-red imagers continue to apply the “slit response function test” to imaging radiometers. From this test, the only figure consistently reported is the milliradian angle at 50 percent of the radiance received by the imager; this is known as the instantaneous field of view (IFOV). The results of the slit response function test enable a user to calculate minimum target size or the distance at which there is a 50 percent probability of detection of a target.
IFOV problems
Applying the slit response function test values poses several problems. Among them are:
By following the "Guideline for Measuring Distance/Target Size Values for Quantitative Thermal Imaging Cameras" published by the Infraspection Institute, the authors were able to calculate spot size ratio values for several modern imaging radiometers for varying percentages of accuracy. See "Table 1 Distance to Target Ratio Values for Imaging Radiometers."
The information contained in Table 1 was derived using the normal lens supplied with each imager. Using telephoto or wide-angle lenses on any imaging radiometer will change ratio values proportionate to the magnification power of the lens. For example, using a 2x telephoto lens will effectively double the ratio values in Table 1; use of a 2x wide-angle lens will effectively halve the listed values.
Several interesting observations were made from the data obtained during testing:
Calculate target size, distance
The distance to spot ratios listed in Table 1 can be used to calculate either the minimum target size or maximum distance at 98 percent accuracy.
To calculate the minimum target size at a given distance, divide the distance from the camera lens to the target by the listed value that corresponds to the shape of the target.
Minimum Target Size = Distance/Listed Value
For example: You want to measure the temperature of a circular target 10 ft away with Imager #3:
Circular Spot Size =10 ft (120 in.)/212 = 0.566 in.
To calculate the maximum distance for a given target size, measure or estimate the size of the target and multiply the listed value by the target size to obtain the maximum distance.
Maximum Distance = Listed Value x Target Size
For example: You are inspecting a 1 in. circular target with Imager #6. How close should you be to measure the temperature?
181 x 1 in. = 181 in. or 15.08 ft
Applying the figures in Table 1, thermographers can develop simple diagrams similar to those commonly used for spot radiometers.
Compare IFOV, spot ratio values
Since many thermographers have long used IFOV values to calculate spot size, it is interesting to note how greatly this method differs from our results. The formula is as follows:
Spot Size = [IFOV (in milliradians)/1000] x Distance to target
Using the published IFOV values for Imager #2 at 30 ft (360 in.) from target, we calculate:
(1.3/1000) x 360 in. = 0.468 in.
Using IFOV value does not consider target shape nor does it state a percentage of accuracy. In fact, this value is derived with the radiometer receiving only 50 percent radiance from the blackbody simulator. IFOV values do not relate to spot size since they are a measurement of individual pixel size. Most imaging radiometers require more than one pixel for accurate temperature measurement. Since detectors vary among manufacturers so will the number and orientation of the pixels required for accurate temperature measurement.
Using the calculated ratio values from Table 1 for Imager #2 at 30 ft (360 in.), we calculate:
Vertical rectangle: 360 in./175 = 2.057 in.
Horizontal rectangle: 360 in/379 = 0.949 in.
Circular: 360 in./144 = 2.500 in.
These spot sizes vary considerably from that calculated by using the IFOV value. As stated earlier, spot size must be smaller than the target in order to accurately measure target temperatures. Radiometric software cannot correct temperatures obtained with an incorrect spot size.
Conclusion
Using IFOV values does not provide meaningful data with respect to spot measurement size of an imaging radiometer. The Infraspection Institute's "Guideline for Measuring Distance/Target Size Values for Quantitative Thermal Imaging Cameras" can be used to calculate distance to spot ratios for any imaging radiometer. This test can be set up with a minimum of equipment and facilities and provides meaningful data about an individual radiometer's accuracy. Knowing a radiometer's accuracy enables a thermographer to better understand his limitations when measuring temperatures.
Thermographers who traditionally have relied on IFOV values should consider using ratio values instead. This is especially true for anyone performing remote infrared inspections of small targets such as overhead power lines or substation components. In these situations, thermographers performing quantitative inspections should consider moving closer to subject targets or using a telephoto lens. MT
This e-mail address is being protected from spambots. You need JavaScript enabled to view it is a staff instructor and This e-mail address is being protected from spambots. You need JavaScript enabled to view it is director of the Infraspection Institute, 425 Ellis St., Burlington, NJ 08016; (609) 386-1281
WIRELESS TEST TOOLS CAN CUT TROUBLESHOOTING TIMEBy: Fluke Corporation
The automation of more and more processes and operations in today’s factories and commercial buildings is helping to reduce energy consumption and increase safety and productivity as never before. However, automation has also added a large dose ofcomplexity for the technicians who maintain and troubleshoot the systems. Click here to learn how Fluke's CNX 3000 Wireless system can help.

New Pig’s PIG® Latching Drum Lids enable quick, easy drum access and secure closure to help meet closed container regulations. Designed to open and close easily with one hand, the Lids keep drum contents dry and pure without hassle.
A partnership with Maintenance Technology and Lubrication Management & Technology keeps your message, products and services in front of 82,000+ decision-makers 24/7/365.
Online, in print and in person, our two synergistic publications are better than one. For more information, click here to contact your MT/LMT Sales Representative today.